Searched for: in-biosketch:true
person:chinil01
Reducing right ventricular pacing burden: algorithms, benefits, and risks
Jankelson, Lior; Bordachar, Pierre; Strik, Mark; Ploux, Sylvain; Chinitz, Larry
Algorithms designed to reduce the burden of right ventricular pacing are widely available in modern implantable pacing devices. To ensure safe and optimal utilization, understanding the properties of these algorithms as well as their possible unfavourable effects is essential. In this review, we discuss in detail the technical and clinical aspects of rhythm management algorithms and update on their significant recent modifications. In addition, we highlight possible adverse phenomena that may be induced by these different pacing algorithms intended to minimize pacing.
PMID: 30428056
ISSN: 1532-2092
CID: 3457302
Behavior of leadless AV synchronous pacing during atrial arrhythmias and stability of the atrial signals over time- Results of the MARVEL Evolve sub-analysis
Garweg, Christophe; Splett, Vincent; Sheldon, Todd J; Chinitz, Larry; Ritter, Philippe; Steinwender, Clemens; Lemme, Francesca; Willems, Rik
INTRODUCTION/BACKGROUND:The MARVEL study demonstrated at a single time point that accelerometer (ACC)-based atrial sensing improves AV synchrony (AVS) in patients with AV block and a Micra pacemaker. The purpose of the MARVEL Evolve sub-study was to assess performance over time. METHODS:This prospective single-center study compared AVS and ACC signals at two visits ≥6 months apart. Custom software was temporarily downloaded into the Micra at each visit and AVS was measured during 30 min at rest. RESULTS:Nine patients from the MARVEL study were enrolled. The mean (±SD) age was 82.3±6.0 years old, 67% were male, and a Micra was implanted for 6.0±6.4 months. High degree AV block was present in 4 patients, whereas 5 with predominantly intrinsic conduction required intermittent pacing for bradycardia. The mean interval between visits was 7.1±0.6 months. Seven patients had normal sinus node function at both visits and were included in a paired analysis. Both ACC signal amplitude (visit 2-visit 1 = 1.4 mG; 95% CI [-25.8 to 28.4 mG]; p = 0.933) and AVS (visit 1: 90.8 %, 95% CI [72.4, 97.4] and visit 2: 91.4%, 95% CI [63.8, 98.5]; p = 0.740) remained stable. Three patients had spontaneous atrial tachycardia. During atrial fibrillation, no atrial contraction was detected or tracked. During atrial flutter, intermittent tracking resulted in a ventricular rate of 60±8 bpm; there was no ventricular pacing > 100 bpm. CONCLUSION/CONCLUSIONS:ACC signals amplitude and performance of AVS pacing were stable over time. During atrial arrhythmias, the AV synchronous pacing mode behaved safely.
PMID: 30687931
ISSN: 1540-8159
CID: 3626312
Simultaneous pace-ablate during CARTO-guided pulmonary vein isolation with a contact-force sensing radiofrequency ablation catheter
Barbhaiya, Chirag R; Aizer, Anthony; Knotts, Robert; Bernstein, Scott; Park, David; Holmes, Douglas; Chinitz, Larry A
PURPOSE/OBJECTIVE:Elimination of pace-capture along pulmonary vein isolation (PVI) lesion sets reduces atrial fibrillation (AF) recurrence in catheter ablation of paroxysmal AF. Pacing from the RF ablation electrode during RF application is prevented within the CARTO electroanatomic mapping system (Biosense Webster, Inc.) due to theoretical safety considerations. We evaluated a method of pacing the distal ablation electrode during RF application in the CARTO system, thus avoiding repeated activation and inactivation of the pacing channel and facilitating immediate recognition of pace-capture loss. We investigated the safety, feasibility, and utility of simultaneous pace-ablate (SPA) during AF ablation with the CARTO-3 system and a contact-force sensing RF ablation catheter. METHODS:Safety of feasibility of SPA was evaluated in 250 patients undergoing first-time AF ablation. Frequency and regional distribution of pace-capture following PVI was evaluated in a cohort of 50 consecutive patients undergoing catheter ablation of paroxysmal AF. RESULTS:SPA was successfully performed in all 250 patients without adverse event. At least one pace-capture site was noted in 22 of 50 PAF patients (44%), and pace-capture following PVI was most common at anterior and superior left atrial sites. There were 2.0 ± 3.3 RF applications during pacing via the distal ablation electrode per patient, and all lesions sets were successfully rendered unexcitable. CONCLUSIONS:Pace-capture along the completed PVI lesion set remains common despite utilization of contact-force sensing RF ablation catheters and automated lesion annotation. Simultaneous pace-ablate in AF ablation using the CARTO system may be safely used to render atrial lesion sets unexcitable.
PMID: 30264289
ISSN: 1572-8595
CID: 3314572
Response to atrial arrhythmias in an atrioventricular synchronous ventricular leadless pacemaker: A case report in a paroxysmal atrial fibrillation patient
Garweg, Christophe; Sheldon, Todd J; Chinitz, Larry; Ritter, Philippe; Steinwender, Clemens; Willems, Rik
PMCID:6301888
PMID: 30581731
ISSN: 2214-0271
CID: 3560032
Echocardiographic Guidance of the Novel WaveCrest Left Atrial Appendage Occlusion Device
Vainrib, Alan F; Bamira, Daniel; Benenstein, Ricardo J; Aizer, Anthony; Chinitz, Larry A; Saric, Muhamed
PMCID:6302034
PMID: 30582095
ISSN: 2468-6441
CID: 3560082
Safety and efficiency of porous-tip contact-force catheter for drug-refractory symptomatic paroxysmal atrial fibrillation ablation: results from the SMART SF trial
Chinitz, Larry A; Melby, Daniel P; Marchlinski, Francis E; Delaughter, Craig; Fishel, Robert S; Monir, George; Patel, Anshul M; Gibson, Douglas N; Athill, Charles A; Ming Boo, Lee; Stagg, Robert; Natale, Andrea
Aims: THERMOCOOL SMARTTOUCH(R) SF Catheter is a new contact-force (CF)-sensing catheter with 56-hole porous tip designed for improved cooling and reduced fluid delivery compared with a standard 6-hole open-irrigated catheter. The SMART SF study examined the periprocedural safety, acute effectiveness, and procedural efficiency of the catheter for drug-refractory symptomatic paroxysmal atrial fibrillation (PAF) ablation. Methods and results: The prospective, open-label, non-randomized SMART-SF was conducted at 17 US sites. Circumferential pulmonary vein (PV) isolation was performed with confirmation of entrance block in all PVs. Stable ablation sites were identified using CARTO VISITAG Module. Primary adverse events (AEs; =1 week of index procedure), periprocedural AEs within 30 days of ablation procedure, acute effectiveness (confirmation of entrance block for targeted PVs), CF, and procedural parameters were assessed. Overall, 165 patients were enrolled (mean age, 62.7 years; male, 57.9%; white, 97%; left ventricular ejection fraction, 60.1 +/- 7%; left atrium diameter, 38.8 +/- 6 mm); 159 underwent radiofrequency ablation and comprised the safety cohort. Primary safety performance criteria were met: primary AE rate was 2.5% (4/159; cardiac tamponade [n = 2], thrombo-embolism [n = 1], transient ischaemic attack [n = 1]). All primary AEs resolved/improved within the 1-month follow-up period. Acute procedural effectiveness was attained in 96.2% (95% confidence interval: 92.0-98.6%) of patients. Procedure time, fluoroscopy time, and fluid delivered were observed in comparison to predecessor catheters. Conclusion: In the SMART-SF trial, the predetermined safety performance goal was met, demonstrating the safety and acute effectiveness of the THERMOCOOL SMARTTOUCH(R) SF Catheter for PAF ablation.
PMID: 29016769
ISSN: 1532-2092
CID: 2732292
REAL-WORLD ECONOMIC AND CLINICAL OUTCOMES FOR PATIENTS WITH ATRIAL FIBRILLATION UNDERGOING RADIOFREQUENCY ABLATION USING THERMOCOOL (R) SF CATHETER VERSUS THERMOCOOL (R) CATHETER [Meeting Abstract]
Chinitz, L.; Goldstein, L. J.; Barnow, A.; Daskiran, M.; Kalsekar, I; Khanna, R.
ISI:000459985602343
ISSN: 1098-3015
CID: 3727582
Implantable Loop Recorder in Inherited Arrhythmia Diseases: A Critical Tool for Symptom Diagnosis and Advanced Risk Stratification [Letter]
Dwivedi, Aeshita; Joza, Jacqueline; Malkani, Kabir; Mendelson, Todd B; Priori, Silvia G; Chinitz, Larry A; Fowler, Steven J; Cerrone, Marina
PMID: 30336885
ISSN: 2405-5018
CID: 3368792
Accelerometer based AV synchronous pacing with a ventricular leadless pacemaker: Results from the Micra AV feasibility studies
Chinitz, Larry; Ritter, Philippe; Khelae, Surinder Kaur; Iacopino, Saverio; Garweg, Christophe; Grazia-Bongiorni, Maria; Neuzil, Petr; Johansen, Jens Brock; Mont, Lluis; Gonzalez, Efrain; Sagi, Venkata; Duray, Gabor Z; Clementy, Nicolas; Sheldon, Todd; Splett, Vincent; Stromberg, Kurt; Wood, Nicole; Steinwender, Clemens
BACKGROUND:Micra is a leadless pacemaker that is implanted in the right ventricle and provides rate response via a 3-axis accelerometer (ACC). Custom software was developed to detect atrial contraction using the ACC enabling AV synchronous pacing. OBJECTIVE:To sense atrial contractions from the Micra ACC signal and provide AV synchronous pacing. METHODS:The MASS/MASS2 early feasibility studies showed intracardiac accelerations related to atrial contraction can be measured via ACC in the Micra leadless pacemaker. MARVEL was a prospective multicenter study designed to characterize the closed-loop performance of an AV synchronous algorithm downloaded into previously implanted Micra devices. AV synchrony (AVS) was measured during 30-minutes of rest and during VVI pacing. AVS was defined as a P-wave visible on a surface ECG followed by a ventricular event <300ms. RESULTS:A total of 64 patients completed the MARVEL study procedure at 12 centers in 9 countries. Patients were implanted with a Micra for a median of 6.0 months (range: 0-41.4). High-degree AV block was present in 33 patients while 31 had predominantly intrinsic conduction during the study. Average AVS during AV algorithm pacing was 87.0% (CI: 81.8%-90.9%); 80.0% in high-degree block patients and 94.4% in patients with intrinsic conduction. AVS was significantly greater (P<0.001) during AV algorithm pacing compared to VVI in high-degree block patients while AVS was maintained in patients with intrinsic conduction. CONCLUSION/CONCLUSIONS:Accelerometer-based atrial sensing is feasible and significantly improves AV synchrony in patients with AV block and a single-chamber leadless pacemaker implanted in the right ventricle.
PMID: 29758405
ISSN: 1556-3871
CID: 3121332
Impact of age on patient selection in leadless pacemaker implant: experience with the Micra transcatheter pacemaker [Meeting Abstract]
Roberts, P. R.; Piccini, J. P.; Clementy, N.; Garweg, C.; Chinitz, L.; Duray, G. Z.; Iacopino, S.; Al Samadi, F.; Ritter, P.; Soejima, K.; Stromberg, K.; Eakley, A. K.; El-Chami, M. F.
ISI:000459824002625
ISSN: 0195-668x
CID: 3727792