Searched for: in-biosketch:yes
person:pc92
Strain rate effects on the mechanical properties and fracture mode of skeletal muscle
Shapiro, Michael; Tovar, Nick; Yoo, Daniel; Sobieraj, Micheal; Gupta, Nikhil; Branski, Ryan C; Coelho, Paulo G
The present study aimed to characterize the mechanical response of beagle sartorius muscle fibers under strain rates that increase logarithmically (0.1mm/min, 1mm/min and 10mm/min), and provide an analysis of the fracture patterns of these tissues via scanning electron microscopy (SEM). Muscle tissue from dogs' sartorius was excised and test specimens were sectioned with a lancet into sections with nominal length, width, and thickness of 7, 2.5 and 0.6mm, respectively. Trimming of the tissue was done so that the loading would be parallel to the direction of the muscle fiber. Samples were immediately tested following excision and failures were observed under the SEM. No statistically significant difference was observed in strength between the 0.1mm/min (2.560+/-0.37MPa) and the 1mm/min (2.702+/-0.55MPa) groups. However, the 10mm/min group (1.545+/-0.50MPa) had a statistically significant lower strength than both the 1mm/min group and the 0.1mm/min group with p<0.01 in both cases. At the 0.1mm/min rate the primary fracture mechanism was that of a shear mode failure of the endomysium with a significant relative motion between fibers. At 1mm/min this continues to be the predominant failure mode. At the 10mm/min strain rate there is a significant change in the fracture pattern relative to other strain rates, where little to no evidence of endomysial shear failure nor of significant motion between fibers was detected.
PMID: 24863204
ISSN: 0928-4931
CID: 1013262
Increased osseointegration effect of BMP-2 on dental implants: An in vivo study
Yoo, Daniel; Tovar, Nick; Jimbo, Ryo; Marin, Charles; Anchieta, Rodolfo B; Machado, Lucas S; Montclare, Jin; Guastaldi, Fernando P S; Janal, Malvin N; Coelho, Paulo G
Application of recombinant human bone morphogenetic protein-2 (rhBMP-2) to implant surfaces has been of great interest due to its osteoinductive potential. However, the optimal coating methodology has not been clarified. The objective of the study was to determine whether the application of rhBMP-2 onto plasma-sprayed hydroxyapatite implant surfaces by immersion in protein solution prior to implant installation would result in significantly improved bone apposition. Using a sheep iliac model, titanium (Ti) and plasma-sprayed calcium phosphate coated (PSCaP) implants uncoated and coated with rhBMP-2 were assessed for their osteogenic effects in the peri-implant area over time in terms of osseointegration and de novo bone formation. After 3 and 6 weeks post-operatively, the samples were retrieved and were subjected to bone-to-implant contact (BIC) and bone area fraction occupancy (BAFO) evaluation. When rhBMP-2 was applied to the PSCaP surface, significant increases in BIC and BAFO were observed at 3 weeks in vivo, whereas when adsorbed directly onto the titanium implant surface, rhBMP-2 did not as effectively improve the bone response (although significantly higher than control Ti). The outcomes of the present study suggested that the combination of plasma-sprayed calcium phosphate surface and rhBMP-2 coating significantly enhanced osseointegration, which validated the postulated hypothesis.
PMID: 23853058
ISSN: 1549-3296
CID: 438662
Patient-specific gingiva-colored abutments: a case series
Sumi, Takashi; Takeshita, Kenji; Takeichi, Takuro; Coelho, Paulo G; Jimbo, Ryo
The purpose of this report is to present a case series of gingiva-colored abutments utilized to meet patients, esthetic needs. The color of the abutment was determined by evaluating the color of the patients, marginal gingiva using a digital shade guide. Thereafter, anodic oxidation was performed on computer-aided design/computer-assisted manufacture abutments. The baseline digital color information was compared with the color of the gingiva after 1.5 years in three patients. The gingival colors as well as the marginal bone and soft tissue levels after 1.5 years were comparable to the baseline values. Thus, the gingiva-colored abutments obtained by anodic oxidation provided improved esthetics, especially for patients with thin gingival biotypes.
PMID: 25006764
ISSN: 0198-7569
CID: 1074352
Mechanical evaluation of four narrow-diameter implant systems
Hirata, Ronaldo; Bonfante, Estevam A; Machado, Lucas S; Tovar, Nick; Coelho, Paulo G
Purpose: This study aimed to evaluate the survival probability of four narrow-diameter implant systems when subjected to fatigue loading. Materials and Methods: Seventy-two narrow-diameter implants to be restored with single-unit crowns were divided into four groups (n = 18): Astra Tech (3.5-mm diameter), with a standard connection (ASC); BioHorizon (3.4-mm diameter), with a standard connection (BSC); Intra-Lock (3.4-mm diameter), with a standard multilobular connection (ISC); and Intra-Lock (3.4-mm diameter), with a modified square connection (IMC). The corresponding abutments were screwed onto the implants, and standardized metal crowns (maxillary central incisors) were cemented and subjected to step-stress accelerated life testing in water. Use-level probability Weibull curves and reliability for 100,000 cycles at 150 and 200 N (90% two-sided confidence intervals) were calculated. Polarized light and scanning electron microscopes were used to access the failure modes. Results: The calculated survival probability for 100,000 cycles at 150 N was approximately 93% in group ASC, 98% in group BSC, 94% in group ISC, and 99% in group IMC. At 200 N, the survival rate was estimated to be approximately < 0.1% for ASC, 77% for BSC, 34% for ISC, and 93% for IMC. Abutment screw fracture was the main failure mode for all groups. Conclusions: Although the probability of survival was not significantly different among systems at a load of 150 N, a significant decrease was observed at 200 N for all groups except IMC.
PMID: 25010880
ISSN: 0893-2174
CID: 1074362
A novel three-dimensional analysis of standardized bone defects by means of confocal scanner and micro-computed tomography
Clozza, Emanuele; Obrecht, Marcel; Dard, Michel; Coelho, Paulo G; Dahlin, Christer; Engebretson, Steven P
OBJECTIVES: The aim of the study was to introduce a novel three-dimensional (3D) method to quantify the relative amount of different tissue components in bone substitute-treated defects by means of integration of confocal laser imaging into micro-computed tomography (muCT) analysis. MATERIALS AND METHODS: One standardized semisaddle intraosseous defect was prepared in the mandibles of six minipigs and scanned by an optical scanner to capture the surface of the fresh defect in a 3D manner. Subsequently, all the defects were filled with a biphasic calcium phosphate material. The animals were divided into two groups of three animals each, which were allowed to heal for 3 and 8 weeks, respectively. muCT analysis followed the two healing periods and was performed on all defect locations. The data from optical scanning and muCT were used for three-dimensional evaluation of bone formation, nonmineralized tissue ratio, and graft degradation. The integration of confocal laser scanning into muCT analysis through a superimposition imaging procedure was conducted using the software Amira (Mercury Computer Systems, Chelmsford, MA, USA). RESULTS: The feasibility of combining the confocal imaging into muCT data with regard to obtaining accurate 3D quantification was demonstrated. The amount of tissue components was identified and quantified in all the investigated samples. Quantitative analysis demonstrated that a significant increase in the amount of bone filling the defect was observed in vivo (p < 0.02) while a significant decrease in the amount of nonmineralized tissue occurred (p < 0.04). No difference in the amount of residual grafting material was detected between 3 and 8 weeks in vivo (p > 0.38). CONCLUSIONS: The combination of confocal imaging and micro-computed tomography techniques allows for analysis of different tissue types over time in vivo. This method has revealed to be a feasible alternative to current bone regeneration quantification methods. CLINICAL RELEVANCE: Assessment of bone formation in a large animal model is a key step in assessing the performance of new bone substitute materials. Reliable and accurate methods are needed for the analysis of the regenerative potential of new materials.
PMID: 23934200
ISSN: 1432-6981
CID: 495272
Effect of abutment screw surface treatment on reliability of implant-supported crowns
Anchieta, Rodolfo Bruniera; Machado, Lucas Silveira; Bonfante, Estevam Augusto; Hirata, Ronaldo; Freitas, Amilcar Chagas Jr; Coelho, Paulo G
Purpose: To evaluate and compare the reliability of implant-supported single crowns cemented onto abutments retained with coated (C) or noncoated (NC) screws and onto platform-switched abutments with coated screws. Materials and Methods: Fifty-four implants (DT Implant 4-mm Standard Platform, Intra-Lock International) were divided into three groups (n = 18 each) as follows: matching-platform abutments secured with noncoated abutment screws (MNC); matching-platform abutments tightened with coated abutment screws (MC); and switched-platform abutments secured with coated abutment screws (SC). Screws were characterized by scanning electron microscopy and x-ray photoelectron spectroscopy (XPS). The specimens were subjected to step-stress accelerated life testing. Use-level probability Weibull curves and reliability for 100,000 cycles at 200 N and 300 N (90% two-sided confidence intervals) were calculated. Polarized light and scanning electron microscopes were used for fractographic analysis. Results: Scanning electron microscopy revealed differences in surface texture; noncoated screws presented the typical machining grooves texture, whereas coated screws presented a plastically deformed surface layer. XPS revealed the same base components for both screws, with the exception of higher degrees of silicon in the SiO2 form for the coated samples. For 100,000 cycles at 300 N, reliability values were 0.06 (0.01 to 0.16), 0.25 (0.09 to 0.45), and 0.25 (0.08 to 0.45), for MNC, MC, and SC, respectively. The most common failure mechanism for MNC was fracture of the abutment screw, followed by bending, or its fracture, along with fracture of the abutment or implant. Coated abutment screws most commonly fractured along with the abutment, irrespective of abutment type. Conclusion: Reliability was higher for both groups with the coated screw than with the uncoated screw. Failure modes differed between coated and uncoated groups.
PMID: 24818196
ISSN: 0882-2786
CID: 979442
The in vivo effect of P-15 coating on early osseointegration
Coelho, Paulo G; Teixeira, Hellen S; Marin, Charles; Witek, Lukasz; Tovar, Nick; Janal, Malvin N; Jimbo, Ryo
The aim of this study was to evaluate mechanically and morphologically the effect of a specific peptide sequence P-15, when incorporated into implant surfaces. Three types of implants were used for the study: Group A: commercially pure titanium implant (blasted and acid etched) + electrochemical thin calcium phosphate deposition, Group B: commercially pure titanium implant (blasted and acid etched) + electrochemical thin calcium phosphate deposition + P-15 incorporation, and as control, Group C: commercially pure titanium implant (blasted and acid etched). After a topographical characterization, transcortical osteotomies were made, and all implant groups (102 implants per group) were randomly placed bilaterally in the tibiae of adult beagle dogs (n = 24). At, 1, 2, and 4 weeks post-surgery, the animals were sacrificed and the samples were retrieved for removal torque tests, for nano indentation, and for histomorphometrical analysis. The results (mean +/- 95% CI) showed that Group B (34.4 +/- 8.7%) presented statistically higher bone-to-implant contact than the other groups (A = 23.9 +/- 7.8%; C = 21.7 +/- 8.3%) at 1 week, indicating an enhanced osteogenesis due to the peptide incorporation. The results suggested that the incorporation of P-15 to implant surfaces increased its bioactivity and the effects were notable especially in the early stages of the healing process. (c) 2013 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2013.
PMID: 24106136
ISSN: 1552-4973
CID: 570362
Evaluation of bone response to various anorganic bovine bone xenografts: an experimental calvaria defect study
Tovar, N; Jimbo, R; Gangolli, R; Perez, L; Manne, L; Yoo, D; Lorenzoni, F; Witek, L; Coelho, P G
This in vivo study investigated the in vivo performance of two newly developed synthetic bone substitutes and compared them to commercially available xenografts (Bio-Oss, Geistlich Pharma AG, Switzerland; OsteoGraf, Dentsply, USA). The materials were tested in a rabbit calvaria model, and the bone forming properties were observed at 4 and 8 weeks after implantation by means of histomorphometry and micro computed tomography (micro-CT). Defects without any graft material were used as negative controls. Micro-CT showed that all materials tested presented new bone formation that filled the defects at both time points, whereas the negative control presented less bone formation, with soft tissue infiltration into the defects. Comparable bone fill percentages were observed for histomorphometric and micro-CT results. Even though no statistically significant difference was found quantitatively between all of the bone graft substitute groups, a higher mean decrease in graft material filling the defects, along with higher remodelling activity, was evident for the experimental materials compared to the commercially available xenografts at 8 weeks. The results indicate that the experimental materials possess high degradability, along with osteoconduction comparable to commercially available xenografts.
PMID: 23948358
ISSN: 0901-5027
CID: 503742
Osteocyte density in the peri-implant bone of implants retrieved after different time periods (4 weeks to 27 years)
Piattelli, Adriano; Artese, Luciano; Penitente, Enrico; Iaculli, Flavia; Degidi, Marco; Mangano, Carlo; Shibli, Jamil Awad; Coelho, Paulo G; Perrotti, Vittoria; Iezzi, Giovanna
Bone tissue is characterized by a constant turnover in response to mechanical stimuli, and osteocytes play an essential role in bone mechanical adaptation. However, little to no information has been published regarding osteocyte density as a function of implantation time in vivo. The aim of this retrospective histological study was to evaluate the osteocyte density of the peri-implant bone in implants retrieved because of different reasons in a time period from 4 weeks to 27 years. A total of 18 samples were included in the present study. Specimens were divided into 3 groups depending on the loading history of the implants: loading between 4 weeks and 7 months (group 1); loading between 1 and 5 years (group 2); loading between 14 and 27 years (group 3). All the samples were histologically evaluated and osteocyte density was obtained using the ratio of the number of osteocytes to the bone-area (mm2 ). The osteocyte density values significantly increased in the Group 2 (1-5 years) compared with Group 1 (4 weeks-7 months), and significantly decreased in the Group 3 (14-27 years) compared to Group 2. No significant differences were detected between Group 1 and Group 3. The decrease in osteocyte density observed in samples that were in vivo for long periods of time under loading is possibly because of the fact that once the bone structure is well aligned and biomechanically competent, a lower number of osteocytes are necessary to keep the tissue homeostasis under loading. (c) 2013 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2013.
PMID: 24106071
ISSN: 1552-4973
CID: 668022
Amoxicillin administrations and its influence on bone repair around osseointegrated implants
Giro, Gabriela; In, Joon; Witek, Lukasz; Granato, Rodrigo; Marin, Charles; Coelho, Paulo G; Ramacciato, Juliana Cama; Lopes Motta, Rogerio Heladio
PURPOSE: The objective of this study was to evaluate the influence of 4 different amoxicillin administration protocols on osseointegration of dental implants. MATERIALS AND METHODS: Thirty-five Wistar rats received an implant in the right tibia and were divided into 5 groups (n = 7): the control group (G1), a group that received a single dose of amoxicillin suspension (40 mg/kg) hour before surgery (G2), a group that received amoxicillin suspension 1 hour before surgery and a 10-mg/kg dose every 12 hours for 3 days (G3), a group that received amoxicillin suspension 1 hour before surgery and a 10-mg/kg dose every 12 hours for 5 days (G4), and a group that received amoxicillin suspension 1 hour before surgery and a 10-mg/kg dose every 12 hours for 7 days (G5). The animals were sacrificed by anesthesia overdose 28 days after implant placement. The samples were retrieved for bone-to-implant contact (BIC) and bone area fraction occupancy (BAFO) analyses. RESULTS: BIC analysis indicated 3 different statistical groups: G1 plus G2, G3, and G4 plus G5. There was no statistical difference between G1 and G2 or between G4 and G5. G3 presented lower values, with statistical difference for G1 plus G2 and G4 plus G5. Also, a statistical difference was found between G1 plus G2 and G4 plus G5. For BAFO evaluation, no statistical difference was found for the experimental groups. CONCLUSION: The results of this study suggest that prolonged use of amoxicillin might have a negative effect on bone formation around implants.
PMID: 24438601
ISSN: 0278-2391
CID: 760262