Effects of Anatomical Differences on Electromagnetic Fields, SAR, and Temperature Change
Alon, Leeor; Deniz, Cem Murat; Carluccio, Giuseppe; Brown, Ryan; Sodickson, Daniel K; Collins, Christopher M
Electromagnetic field simulations are increasingly used to assure RF safety of patients during MRI exams. In practice, however, tissue property distribution of the patient being imaged is not known, but may be represented with a pre-existing model. Repeatedly, agreement in transmit magnetic (B1 +) field distributions between two geometries has been used to suggest agreement in heating distributions. Here we examine relative effects of anatomical differences on B1 + distribution, Specific Absorption Rate (SAR) and temperature change (DeltaT). Numerical simulations were performed for a single surface coil positioned adjacent a homogeneous phantom and bovine phantom, each with slight geometric variations, and adjacent two different human body models. Experimental demonstration was performed on a bovine phantom using MR thermometry and B1 + mapping. Simulations and experiments demonstrate that B1 + distributions in different samples can be well correlated, while notable difference in maximum SAR and DeltaT occur. This work illustrates challenges associated with utilizing simulations or experiments for RF safety assurance purposes. Reliance on B1 + distributions alone for validation of simulations and/or experiments with a sample or subject for assurance of safety in another should be performed with caution.
PMCID:4847547
PMID: 27134586
ISSN: 1552-5031
CID: 2101092
High-performance radiofrequency coils for Na MRI: brain and musculoskeletal applications
Wiggins, Graham C; Brown, Ryan; Lakshmanan, Karthik
23 Na RF coil design for brain and MSK applications presents a number of challenges, including poor coil loading for arrays of small coils and SNR penalties associated with providing 1 H capability with the same coil. The basics of RF coil design are described, as well as a review of historical approaches to dual tuning. There follows a review of published high performance coil designs for MSK and brain imaging. Several coil designs have been demonstrated at 7T and 3T which incorporate close-fitting receive arrays and in some cases design features which provide 1 H imaging with little penalty to 23 Na sensitivity. The 'nested coplanar loop' approach is examined, in which small transmit-receive 1 H elements are placed within each 23 Na loop, presenting only a small perturbation to 23 Na performance and minimizing RF shielding issues. Other designs incorporating transmit-receive arrays for 23 Na and 1 H are discussed including a 9.4 T 23 Na / 1 H brain coil. Great gains in 23 Na SNR have been made with many of these designs, but simultaneously achieving high performance for 1H remains elusive
PMCID:4713340
PMID: 26404631
ISSN: 1099-1492
CID: 1787022
Dependence of B1+ and B1- Field Patterns of Surface Coils on the Electrical Properties of the Sample and the MR Operating Frequency
Vaidya, Manushka V; Collins, Christopher M; Sodickson, Daniel K; Brown, Ryan; Wiggins, Graham C; Lattanzi, Riccardo
In high field MRI, the spatial distribution of the radiofrequency magnetic ( B1) field is usually affected by the presence of the sample. For hardware design and to aid interpretation of experimental results, it is important both to anticipate and to accurately simulate the behavior of these fields. Fields generated by a radiofrequency surface coil were simulated using dyadic Green's functions, or experimentally measured over a range of frequencies inside an object whose electrical properties were varied to illustrate a variety of transmit [Formula: see text] and receive [Formula: see text] field patterns. In this work, we examine how changes in polarization of the field and interference of propagating waves in an object can affect the B1 spatial distribution. Results are explained conceptually using Maxwell's equations and intuitive illustrations. We demonstrate that the electrical conductivity alters the spatial distribution of distinct polarized components of the field, causing "twisted" transmit and receive field patterns, and asymmetries between [Formula: see text] and [Formula: see text]. Additionally, interference patterns due to wavelength effects are observed at high field in samples with high relative permittivity and near-zero conductivity, but are not present in lossy samples due to the attenuation of propagating EM fields. This work provides a conceptual framework for understanding B1 spatial distributions for surface coils and can provide guidance for RF engineers.
PMCID:5082994
PMID: 27795697
ISSN: 1552-5031
CID: 2296462
A nested phosphorus and proton coil array for brain magnetic resonance imaging and spectroscopy
Brown, Ryan; Lakshmanan, Karthik; Madelin, Guillaume; Parasoglou, Prodromos
A dual-nuclei radiofrequency coil array was constructed for phosphorus and proton magnetic resonance imaging and spectroscopy of the human brain at 7T. An eight-channel transceive degenerate birdcage phosphorus module was implemented to provide whole-brain coverage and significant sensitivity improvement over a standard dual-tuned loop coil. A nested eight-channel proton module provided adequate sensitivity for anatomical localization without substantially sacrificing performance on the phosphorus module. The developed array enabled phosphorus spectroscopy, a saturation transfer technique to calculate the global creatine kinase forward reaction rate, and single-metabolite whole-brain imaging with 1.4cm nominal isotropic resolution in 15min (2.3cm actual resolution), while additionally enabling 1mm isotropic proton imaging. This study demonstrates that a multi-channel array can be utilized for phosphorus and proton applications with improved coverage and/or sensitivity over traditional single-channel coils. The efficient multi-channel coil array, time-efficient pulse sequences, and the enhanced signal strength available at ultra-high fields can be combined to allow volumetric assessment of the brain and could provide new insights into the underlying energy metabolism impairment in several neurodegenerative conditions, such as Alzheimer's and Parkinson's diseases, as well as mental disorders such as schizophrenia.
PMCID:4651763
PMID: 26375209
ISSN: 1095-9572
CID: 1779282
Response [Comment]
Chang, Gregory; Brown, Ryan; Regatte, Ravinder R; Rajapakse, Chamith S
PMID: 25785289
ISSN: 1527-1315
CID: 4069922
Characterization of a dielectric phantom for high-field magnetic resonance imaging applications
Duan, Qi; Duyn, Jeff H; Gudino, Natalia; de Zwart, Jacco A; van Gelderen, Peter; Sodickson, Daniel K; Brown, Ryan
PURPOSE: In this work, a generic recipe for an inexpensive and nontoxic phantom was developed within a range of biologically relevant dielectric properties from 150 MHz to 4.5 GHz. METHODS: The recipe includes deionized water as the solvent, NaCl to primarily control conductivity, sucrose to primarily control permittivity, agar-agar to gel the solution and reduce heat diffusivity, and benzoic acid to preserve the gel. Two hundred and seventeen samples were prepared to cover the feasible range of NaCl and sucrose concentrations. Their dielectric properties were measured using a commercial dielectric probe and were fitted to a 3D polynomial to generate a recipe describing the properties as a function of NaCl concentration, sucrose concentration, and frequency. RESULTS: Results indicated that the intuitive linear and independent relationships between NaCl and conductivity and between sucrose and permittivity are not valid. A generic polynomial recipe was developed to characterize the complex relationship between the solutes and the resulting dielectric values and has been made publicly available as a web application. In representative mixtures developed to mimic brain and muscle tissue, less than 2% difference was observed between the predicted and measured conductivity and permittivity values. CONCLUSIONS: It is expected that the recipe will be useful for generating dielectric phantoms for general magnetic resonance imaging (MRI) coil development at high magnetic field strength, including coil safety evaluation as well as pulse sequence evaluation (including B1 (+) mapping, B1 (+) shimming, and selective excitation pulse design), and other non-MRI applications which require biologically equivalent dielectric properties.
PMCID:4281072
PMID: 25281973
ISSN: 0094-2405
CID: 1299002
Finite Element Analysis Applied to 3-T MR Imaging of Proximal Femur Microarchitecture: Lower Bone Strength in Patients with Fragility Fractures Compared with Control Subjects
Chang, Gregory; Honig, Stephen; Brown, Ryan; Deniz, Cem M; Egol, Kenneth A; Babb, James S; Regatte, Ravinder R; Rajapakse, Chamith S
Purpose To determine the feasibility of using finite element analysis applied to 3-T magnetic resonance (MR) images of proximal femur microarchitecture for detection of lower bone strength in subjects with fragility fractures compared with control subjects without fractures. Materials and Methods This prospective study was institutional review board approved and HIPAA compliant. Written informed consent was obtained. Postmenopausal women with (n = 22) and without (n = 22) fragility fractures were matched for age and body mass index. All subjects underwent standard dual-energy x-ray absorptiometry. Images of proximal femur microarchitecture were obtained by using a high-spatial-resolution three-dimensional fast low-angle shot sequence at 3 T. Finite element analysis was applied to compute elastic modulus as a measure of strength in the femoral head and neck, Ward triangle, greater trochanter, and intertrochanteric region. The Mann-Whitney test was used to compare bone mineral density T scores and elastic moduli between the groups. The relationship (R2) between elastic moduli and bone mineral density T scores was assessed. Results Patients with fractures showed lower elastic modulus than did control subjects in all proximal femur regions (femoral head, 8.51-8.73 GPa vs 9.32-9.67 GPa; P = .04; femoral neck, 3.11-3.72 GPa vs 4.39-4.82 GPa; P = .04; Ward triangle, 1.85-2.21 GPa vs 3.98-4.13 GPa; P = .04; intertrochanteric region, 1.62-2.18 GPa vs 3.86-4.47 GPa; P = .006-.007; greater trochanter, 0.65-1.21 GPa vs 1.96-2.62 GPa; P = .01-.02), but no differences in bone mineral density T scores. There were weak relationships between elastic moduli and bone mineral density T scores in patients with fractures (R2 = 0.25-0.31, P = .02-.04), but not in control subjects. Conclusion Finite element analysis applied to high-spatial-resolution 3-T MR images of proximal femur microarchitecture can allow detection of lower elastic modulus, a marker of bone strength, in subjects with fragility fractures compared with control subjects. MR assessment of proximal femur strength may provide information about bone quality that is not provided by dual-energy x-ray absorptiometry. (c) RSNA, 2014.
PMCID:4263634
PMID: 24689884
ISSN: 0033-8419
CID: 895822
Feasibility of three-dimensional MRI of proximal femur microarchitecture at 3 tesla using 26 receive elements without and with parallel imaging
Chang, Gregory; Deniz, Cem M; Honig, Stephen; Rajapakse, Chamith S; Egol, Kenneth; Regatte, Ravinder R; Brown, Ryan
PURPOSE: High-resolution imaging of deeper anatomy such as the hip is challenging due to low signal-to-noise ratio (SNR), necessitating long scan times. Multi-element coils can increase SNR and reduce scan time through parallel imaging (PI). We assessed the feasibility of using a 26-element receive coil setup to perform 3 Tesla (T) MRI of proximal femur microarchitecture without and with PI. MATERIALS AND METHODS: This study had institutional review board approval. We scanned 13 subjects on a 3T scanner using 26 receive-elements and a three-dimensional fast low-angle shot (FLASH) sequence without and with PI (acceleration factors [AF] 2, 3, 4). We assessed SNR, depiction of individual trabeculae, PI performance (1/g-factor), and image quality with PI (1 = nonvisualization to 5 = excellent). RESULTS: SNR maps demonstrate higher SNR for the 26-element setup compared with a 12-element setup for hip MRI. Without PI, individual proximal femur trabeculae were well-depicted, including microarchitectural deterioration in osteoporotic subjects. With PI, 1/g values for the 26-element/12-element receive-setup were 0.71/0.45, 0.56/0.25, and 0.44/0.08 at AF2, AF3, and AF4, respectively. Image quality was: AF1, excellent (4.8 +/- 0.4); AF2, good (4.2 +/- 1.0); AF3, average (3.3 +/- 1.0); AF4, nonvisualization (1.4 +/- 0.9). CONCLUSION: A 26-element receive-setup permits 3T MRI of proximal femur microarchitecture with good image quality up to PI AF2. J. Magn. Reson. Imaging 2014;40:229-238. (c) 2013 Wiley Periodicals, Inc.
PMCID:4004721
PMID: 24711013
ISSN: 1053-1807
CID: 1042112
MRI of the hip at 7T: Feasibility of bone microarchitecture, high-resolution cartilage, and clinical imaging
Chang, Gregory; Deniz, Cem M; Honig, Stephen; Egol, Kenneth; Regatte, Ravinder R; Zhu, Yudong; Sodickson, Daniel K; Brown, Ryan
PURPOSE: To demonstrate the feasibility of performing bone microarchitecture, high-resolution cartilage, and clinical imaging of the hip at 7T. MATERIALS AND METHODS: This study had Institutional Review Board approval. Using an 8-channel coil constructed in-house, we imaged the hips of 15 subjects on a 7T magnetic resonance imaging (MRI) scanner. We applied: 1) a T1-weighted 3D fast low angle shot (3D FLASH) sequence (0.23 x 0.23 x 1-1.5 mm3 ) for bone microarchitecture imaging; 2) T1-weighted 3D FLASH (water excitation) and volumetric interpolated breath-hold examination (VIBE) sequences (0.23 x 0.23 x 1.5 mm3 ) with saturation or inversion recovery-based fat suppression for cartilage imaging; 3) 2D intermediate-weighted fast spin-echo (FSE) sequences without and with fat saturation (0.27 x 0.27 x 2 mm) for clinical imaging. RESULTS: Bone microarchitecture images allowed visualization of individual trabeculae within the proximal femur. Cartilage was well visualized and fat was well suppressed on FLASH and VIBE sequences. FSE sequences allowed visualization of cartilage, the labrum (including cartilage and labral pathology), joint capsule, and tendons. CONCLUSION: This is the first study to demonstrate the feasibility of performing a clinically comprehensive hip MRI protocol at 7T, including high-resolution imaging of bone microarchitecture and cartilage, as well as clinical imaging. J. Magn. Reson. Imaging 2013;. (c) 2013 Wiley Periodicals, Inc.
PMCID:3962810
PMID: 24115554
ISSN: 1053-1807
CID: 571382
Breast MRI at 7 Tesla with a bilateral coil and robust fat suppression
Brown, Ryan; Storey, Pippa; Geppert, Christian; McGorty, Kellyanne; Klautau Leite, Ana Paula; Babb, James; Sodickson, Daniel K; Wiggins, Graham C; Moy, Linda
PURPOSE: To develop a bilateral coil and fat suppressed T1-weighted sequence for 7 Tesla (T) breast MRI. MATERIALS AND METHODS: A dual-solenoid coil and three-dimensional (3D) T1w gradient echo sequence with B1 + insensitive fat suppression (FS) were developed. T1w FS image quality was characterized through image uniformity and fat-water contrast measurements in 11 subjects. Signal-to-noise ratio (SNR) and flip angle maps were acquired to assess the coil performance. Bilateral contrast-enhanced and unilateral high resolution (0.6 mm isotropic, 6.5 min acquisition time) imaging highlighted the 7T SNR advantage. RESULTS: Reliable and effective FS and high image quality was observed in all subjects at 7T, indicating that the custom coil and pulse sequence were insensitive to high-field obstacles such as variable tissue loading. 7T and 3T image uniformity was similar (P = 0.24), indicating adequate 7T B1 + uniformity. High 7T SNR and fat-water contrast enabled 0.6 mm isotropic imaging and visualization of a high level of fibroglandular tissue detail. CONCLUSION: 7T T1w FS bilateral breast imaging is feasible with a custom radiofrequency (RF) coil and pulse sequence. Similar image uniformity was achieved at 7T and 3T, despite different RF field behavior and variable coil-tissue interaction due to anatomic differences that might be expected to alter magnetic field patterns. J. Magn. Reson. Imaging 2013. (c) 2013 Wiley Periodicals, Inc.
PMCID:3945054
PMID: 24123517
ISSN: 1053-1807
CID: 571402