Try a new search

Format these results:

Searched for:

person:jtm13

Total Results:

263


Programmable Bio-nanochip Platform: A Point-of-Care Biosensor System with the Capacity To Learn

McRae, Michael P; Simmons, Glennon; Wong, Jorge; McDevitt, John T
The combination of point-of-care (POC) medical microdevices and machine learning has the potential transform the practice of medicine. In this area, scalable lab-on-a-chip (LOC) devices have many advantages over standard laboratory methods, including faster analysis, reduced cost, lower power consumption, and higher levels of integration and automation. Despite significant advances in LOC technologies over the years, several remaining obstacles are preventing clinical implementation and market penetration of these novel medical microdevices. Similarly, while machine learning has seen explosive growth in recent years and promises to shift the practice of medicine toward data-intensive and evidence-based decision making, its uptake has been hindered due to the lack of integration between clinical measurements and disease determinations. In this Account, we describe recent developments in the programmable bio-nanochip (p-BNC) system, a biosensor platform with the capacity for learning. The p-BNC is a "platform to digitize biology" in which small quantities of patient sample generate immunofluorescent signal on agarose bead sensors that is optically extracted and converted to antigen concentrations. The platform comprises disposable microfluidic cartridges, a portable analyzer, automated data analysis software, and intuitive mobile health interfaces. The single-use cartridges are fully integrated, self-contained microfluidic devices containing aqueous buffers conveniently embedded for POC use. A novel fluid delivery method was developed to provide accurate and repeatable flow rates via actuation of the cartridge's blister packs. A portable analyzer instrument was designed to integrate fluid delivery, optical detection, image analysis, and user interface, representing a universal system for acquiring, processing, and managing clinical data while overcoming many of the challenges facing the widespread clinical adoption of LOC technologies. We demonstrate the p-BNC's flexibility through the completion of multiplex assays within the single-use disposable cartridges for three clinical applications: prostate cancer, ovarian cancer, and acute myocardial infarction. Toward the goal of creating "sensors that learn", we have developed and describe here the Cardiac ScoreCard, a clinical decision support system for a spectrum of cardiovascular disease. The Cardiac ScoreCard approach comprises a comprehensive biomarker panel and risk factor information in a predictive model capable of assessing early risk and late-stage disease progression for heart attack and heart failure patients. These marker-driven tests have the potential to radically reduce costs, decrease wait times, and introduce new options for patients needing regular health monitoring. Further, these efforts demonstrate the clinical utility of fusing data from information-rich biomarkers and the Internet of Things (IoT) using predictive analytics to generate single-index assessments for wellness/illness status. By promoting disease prevention and personalized wellness management, tools of this nature have the potential to improve health care exponentially.
PMID: 27380817
ISSN: 1520-4898
CID: 2175992

Cardiac ScoreCard: A Diagnostic Multivariate Index Assay System for Predicting a Spectrum of Cardiovascular Disease

McRae, Michael P; Bozkurt, Biykem; Ballantyne, Christie M; Sanchez, Ximena; Christodoulides, Nicolaos; Simmons, Glennon; Nambi, Vijay; Misra, Arunima; Miller, Craig S; Ebersole, Jeffrey L; Campbell, Charles; McDevitt, John T
Clinical decision support systems (CDSSs) have the potential to save lives and reduce unnecessary costs through early detection and frequent monitoring of both traditional risk factors and novel biomarkers for cardiovascular disease (CVD). However, the widespread adoption of CDSSs for the identification of heart diseases has been limited, likely due to the poor interpretability of clinically relevant results and the lack of seamless integration between measurements and disease predictions. In this paper we present the Cardiac ScoreCard-a multivariate index assay system with the potential to assist in the diagnosis and prognosis of a spectrum of CVD. The Cardiac ScoreCard system is based on lasso logistic regression techniques which utilize both patient demographics and novel biomarker data for the prediction of heart failure (HF) and cardiac wellness. Lasso logistic regression models were trained on a merged clinical dataset comprising 579 patients with 6 traditional risk factors and 14 biomarker measurements. The prediction performance of the Cardiac ScoreCard was assessed with 5-fold cross-validation and compared with reference methods. The experimental results reveal that the ScoreCard models improved performance in discriminating disease versus non-case (AUC = 0.8403 and 0.9412 for cardiac wellness and HF, respectively), and the models exhibit good calibration. Clinical insights to the prediction of HF and cardiac wellness are provided in the form of logistic regression coefficients which suggest that augmenting the traditional risk factors with a multimarker panel spanning a diverse cardiovascular pathophysiology provides improved performance over reference methods. Additionally, a framework is provided for seamless integration with biomarker measurements from point-of-care medical microdevices, and a lasso-based feature selection process is described for the down-selection of biomarkers in multimarker panels.
PMCID:6715313
PMID: 31467464
ISSN: 0957-4174
CID: 4066542

Challenges and opportunities for translating medical microdevices: insights from the programmable bio-nano-chip

McRae, Michael P; Simmons, Glennon; McDevitt, John T
This perspective highlights the major challenges for the bioanalytical community, in particular the area of lab-on-a-chip sensors, as they relate to point-of-care diagnostics. There is a strong need for general-purpose and universal biosensing platforms that can perform multiplexed and multiclass assays on real-world clinical samples. However, the adoption of novel lab-on-a-chip/microfluidic devices has been slow as several key challenges remain for the translation of these new devices to clinical practice. A pipeline of promising medical microdevice technologies will be made possible by addressing the challenges of integration, failure to compete with cost and performance of existing technologies, requisite for new content, and regulatory approval and clinical adoption.
PMCID:4870725
PMID: 27071710
ISSN: 1757-6199
CID: 3110592

Point-of-Care Technologies for Precision Cardiovascular Care and Clinical Research: National Heart, Lung, and Blood Institute Working Group

King, Kevin; Grazette, Luanda P; Paltoo, Dina N; McDevitt, John T; Sia, Samuel K; Barrett, Paddy M; Apple, Fred S; Gurbel, Paul A; Weissleder, Ralph; Leeds, Hilary; Iturriaga, Erin J; Rao, Anupama; Adhikari, Bishow; Desvigne-Nickens, Patrice; Galis, Zorina S; Libby, Peter
Point-of-care technologies (POC or POCT) are enabling innovative cardiovascular diagnostics that promise to improve patient care across diverse clinical settings. The National Heart, Lung, and Blood Institute convened a working group to discuss POCT in cardiovascular medicine. The multidisciplinary working group, which included clinicians, scientists, engineers, device manufacturers, regulatory officials, and program staff, reviewed the state of the POCT field; discussed opportunities for POCT to improve cardiovascular care, realize the promise of precision medicine, and advance the clinical research enterprise; and identified barriers facing translation and integration of POCT with existing clinical systems. A POCT development roadmap emerged to guide multidisciplinary teams of biomarker scientists, technologists, health care providers, and clinical trialists as they: 1) formulate needs assessments; 2) define device design specifications; 3) develop component technologies and integrated systems; 4) perform iterative pilot testing; and 5) conduct rigorous prospective clinical testing to ensure that POCT solutions have substantial effects on cardiovascular care.
PMCID:4787294
PMID: 26977455
ISSN: 2452-302x
CID: 2047162

Next Generation Programmable Bio-Nano-Chip System for On-Site Detection in Oral Fluids

Christodoulides, Nicolaos; De La Garza, Richard 2nd; Simmons, Glennon W; McRae, Michael P; Wong, Jorge; Newton, Thomas F; Kosten, Thomas R; Haque, Ahmed; McDevitt, John T
Current on-site drug of abuse detection methods involve invasive sampling of blood and urine specimens, or collection of oral fluid, followed by qualitative screening tests using immunochromatographic cartridges. Test confirmation and quantitative assessment of a presumptive positive are then provided by remote laboratories, an inefficient and costly process decoupled from the initial sampling. Recently, a new noninvasive oral fluid sampling approach that is integrated with the chip-based Programmable Bio-Nano-Chip (p-BNC) platform has been developed for the rapid (~ 10 minutes), sensitive detection (~ ng/ml) and quantitation of 12 drugs of abuse. Furthermore, the system can provide the time-course of select drug and metabolite profiles in oral fluids. For cocaine, we observed three slope components were correlated with cocaine-induced impairment using this chip-based p-BNC detection modality. Thus, this p-BNC has significant potential for roadside drug testing by law enforcement officers. Initial work reported on chip-based drug detection was completed using 'macro' or "chip in the lab" prototypes, that included metal encased "flow cells", external peristaltic pumps and a bench-top analyzer system instrumentation. We now describe the next generation miniaturized analyzer instrumentation along with customized disposables and sampling devices. These tools will offer real-time oral fluid drug monitoring capabilities, to be used for roadside drug testing as well as testing in clinical settings as a non-invasive, quantitative, accurate and sensitive tool to verify patient adherence to treatment.
PMCID:4765139
PMID: 26925466
ISSN: n/a
CID: 2046142

Programmable bio-nano-chip system: a flexible point-of-care platform for bioscience and clinical measurements

McRae, Michael P; Simmons, Glennon W; Wong, Jorge; Shadfan, Basil; Gopalkrishnan, Sanjiv; Christodoulides, Nicolaos; McDevitt, John T
The development of integrated instrumentation for universal bioassay systems serves as a key goal for the lab-on-a-chip community. The programmable bio-nano-chip (p-BNC) system is a versatile multiplexed and multiclass chemical- and bio-sensing system for bioscience and clinical measurements. The system is comprised of two main components, a disposable cartridge and a portable analyzer. The customizable single-use plastic cartridges, which now can be manufactured in high volumes using injection molding, are designed for analytical performance, ease of use, reproducibility, and low cost. These labcard devices implement high surface area nano-structured biomarker capture elements that enable high performance signaling and are index-matched to real-world biological specimens. This detection modality, along with the convenience of on-chip fluid storage in blisters and self-contained waste, represents a standard process to digitize biological signatures at the point-of-care. A companion portable analyzer prototype has been developed to integrate fluid motivation, optical detection, and automated data analysis, and it serves as the human interface for complete assay automation. In this report, we provide a systems-level perspective of the p-BNC universal biosensing platform with an emphasis on flow control, device integration, and automation. To demonstrate the flexibility of the p-BNC, we distinguish diseased and non-case patients across three significant disease applications: prostate cancer, ovarian cancer, and acute myocardial infarction. Progress towards developing a rapid 7 minute myoglobin assay is presented using the fully automated p-BNC system.
PMCID:4589532
PMID: 26308851
ISSN: 1473-0189
CID: 1745622

Interobserver agreement in dysplasia grading: toward an enhanced gold standard for clinical pathology trials

Speight, Paul M; Abram, Timothy J; Floriano, Pierre N; James, Robert; Vick, Julie; Thornhill, Martin H; Murdoch, Craig; Freeman, Christine; Hegarty, Anne M; D'Apice, Katy; Kerr, A Ross; Phelan, Joan; Corby, Patricia; Khouly, Ismael; Vigneswaran, Nadarajah; Bouquot, Jerry; Demian, Nagi M; Weinstock, Y Etan; Redding, Spencer W; Rowan, Stephanie; Yeh, Chih-Ko; McGuff, H Stan; Miller, Frank R; McDevitt, John T
OBJECTIVE: Interobserver agreement in the context of oral epithelial dysplasia (OED) grading has been notoriously unreliable and can impose barriers for developing new molecular markers and diagnostic technologies. This paper aimed to report the details of a 3-stage histopathology review and adjudication process with the goal of achieving a consensus histopathologic diagnosis of each biopsy. STUDY DESIGN: Two adjacent serial histologic sections of oral lesions from 846 patients were independently scored by 2 different pathologists from a pool of 4. In instances where the original 2 pathologists disagreed, a third, independent adjudicating pathologist conducted a review of both sections. If a majority agreement was not achieved, the third stage involved a face-to-face consensus review. RESULTS: Individual pathologist pair kappa values ranged from 0.251 to 0.706 (fair-good) before the 3-stage review process. During the initial review phase, the 2 pathologists agreed on a diagnosis for 69.9% of the cases. After the adjudication review by a third pathologist, an additional 22.8% of cases were given a consensus diagnosis (agreement of 2 out of 3 pathologists). After the face-to-face review, the remaining 7.3% of cases had a consensus diagnosis. CONCLUSIONS: The use of the defined protocol resulted in a substantial increase (30%) in diagnostic agreement and has the potential to improve the level of agreement for establishing gold standards for studies based on histopathologic diagnosis.
PMCID:4564355
PMID: 26216170
ISSN: 2212-4411
CID: 1698802

Application of programmable bio-nano-chip system for the quantitative detection of drugs of abuse in oral fluids

Christodoulides, Nicolaos; De La Garza Ii, Richard; Simmons, Glennon W; McRae, Michael P; Wong, Jorge; Newton, Thomas F; Smith, Regina; Mahoney Iii, James J; Hohenstein, Justin; Gomez, Sobeyda; Floriano, Pierre N; Talavera, Humberto; Sloan, Daniel J; Moody, David E; Andrenyak, David M; Kosten, Thomas R; Haque, Ahmed; McDevitt, John T
OBJECTIVE: There is currently a gap in on-site drug of abuse monitoring. Current detection methods involve invasive sampling of blood and urine specimens, or collection of oral fluid, followed by qualitative screening tests using immunochromatographic cartridges. While remote laboratories then may provide confirmation and quantitative assessment of a presumptive positive, this instrumentation is expensive and decoupled from the initial sampling making the current drug-screening program inefficient and costly. The authors applied a noninvasive oral fluid sampling approach integrated with the in-development chip-based Programmable bio-nano-chip (p-BNC) platform for the detection of drugs of abuse. METHOD: The p-BNC assay methodology was applied for the detection of tetrahydrocannabinol, morphine, amphetamine, methamphetamine, cocaine, methadone and benzodiazepines, initially using spiked buffered samples and, ultimately, using oral fluid specimen collected from consented volunteers. RESULTS: Rapid ( approximately 10min), sensitive detection ( approximately ng/mL) and quantitation of 12 drugs of abuse was demonstrated on the p-BNC platform. Furthermore, the system provided visibility to time-course of select drug and metabolite profiles in oral fluids; for the drug cocaine, three regions of slope were observed that, when combined with concentration measurements from this and prior impairment studies, information about cocaine-induced impairment may be revealed. CONCLUSIONS: This chip-based p-BNC detection modality has significant potential to be used in the future by law enforcement officers for roadside drug testing and to serve a variety of other settings, including outpatient and inpatient drug rehabilitation centers, emergency rooms, prisons, schools, and in the workplace.
PMCID:4509839
PMID: 26048639
ISSN: 1879-0046
CID: 1615882

Programmable bio-nanochip-based cytologic testing of oral potentially malignant disorders in Fanconi anemia

Floriano, P N; Abram, T; Taylor, L; Le, C; Talavera, H; Nguyen, M; Raja, R; Gillenwater, A; McDevitt, J; Vigneswaran, N
Fanconi anemia (FA) is caused by mutations of DNA repair genes. The risk of oral squamous cell carcinoma (OSCC) among FA patients is 800-folds higher than in the general population. Early detection of OSCC, preferably at it precursor stage, is critical in FA patients to improve their survival. In an ongoing clinical trial, we are evaluating the effectiveness of the programmable bio-nanochip (p-BNC)-based oral cytology test in diagnosing oral potentially malignant disorders (OPMD) in non-FA patients. We used this test to compare cytomorphometric and molecular biomarkers in OSCC cell lines derived from FA and non-FA patients to brush biopsy samples of a FA patient with OPMD and normal mucosa of healthy volunteers. Our data showed that expression patterns of molecular biomarkers were not notably different between sporadic and FA-OSCC cell lines. The p-BNC assay revealed significant differences in cytometric parameters and biomarker MCM2 expression between cytobrush samples of the FA patient and cytobrush samples of normal oral mucosa obtained from healthy volunteers. Microscopic examination of the FA patient's OPMD confirmed the presence of dysplasia. Our pilot data suggests that the p-BNC brush biopsy test recognized dysplastic oral epithelial cells in a brush biopsy sample of a FA patient.
PMCID:4457567
PMID: 25662766
ISSN: 1601-0825
CID: 2911682

A multiplexable, microfluidic platform for the rapid quantitation of a biomarker panel for early ovarian cancer detection at the point-of-care

Shadfan, Basil H; Simmons, Archana R; Simmons, Glennon W; Ho, Andy; Wong, Jorge; Lu, Karen; Bast, Robert C Jr; McDevitt, John T
Point-of-care (POC) diagnostic platforms have the potential to enable low-cost, large-scale screening. As no single biomarker is shed by all ovarian cancers, multiplexed biomarker panels promise improved sensitivity and specificity to address the unmet need for early detection of ovarian cancer. We have configured the programmable bio-nano-chip (p-BNC) - a multiplexable, microfluidic, modular platform - to quantify a novel multimarker panel comprising CA125, HE4, MMP-7 and CA72-4. The p-BNC is a bead-based immunoanalyzer system with a credit-card-sized footprint that integrates automated sample metering, bubble and debris removal, reagent storage and waste disposal, permitting POC analysis. Multiplexed p-BNC immunoassays demonstrated high specificity, low cross-reactivity, low limits of detection suitable for early detection, and a short analysis time of 43 minutes. Day-to-day variability, a critical factor for longitudinally monitoring biomarkers, ranged between 5.4-10.5%, well below the biological variation for all four markers. Biomarker concentrations for 31 late-stage sera correlated well (R2 = 0.71 to 0.93 for various biomarkers) with values obtained on the Luminex(R) platform. In a 31 patient cohort encompassing early- and late-stage ovarian cancers along with benign and healthy controls, the multiplexed p-BNC panel was able to distinguish cases from controls with 68.7% sensitivity at 80% specificity. Utility for longitudinal biomarker monitoring was demonstrated with pre-diagnostic sera from 2 cases and 4 controls. Taken together, the p-BNC shows strong promise as a diagnostic tool for large-scale screening that takes advantage of faster results and lower costs while leveraging possible improvement in sensitivity and specificity from biomarker panels.
PMCID:4465398
PMID: 25388014
ISSN: 1940-6215
CID: 1390382